Specifications
Base Hardware
- Industrialized rack-mounted computer chassis
- IBM PC-AT compatible 66 MHz 486 DX2 CPU card with 4 MB RAM,
floppy and hard disk controllers, two serial ports, printer port, and VGA
- 320 MB hard disk and 1.44-MB floppy (others available)
- Smart 150 KHz A/D card
- 16-channel analog output card with 24 bits digital I/O
- Timer-counter board
- Rack-mounted VGA monitor
- Industrial keyboard
- 3 KVA UPS (Uninterruptible Power Supply)
- Optional cabinet air conditioner
- Optional intelligent Ethernet interface card
|
Operator Interface
- Function keys
- Test selection and initialization menu defined by user
- Test Header (identification) window (user defined)
- Operator comments are time/date stamped and written to a log
file
- User defined operator reminder messages can be generated by the
test sequence
- User defined operator acknowledge messages can hold test
sequence until the operator responds
- User defined manual entry messages prompt the operator to type
in manual data
- Alarm messages are written to screen and log file
- Calibration information logged to audit file
|
Inputs and Outputs
- 64 channels of analog input
- Each channel has its own transformer isolated signal
conditioning
- 24 thermocouple channels (type J, K, or T)
- 24 pressure channels (0.5 to 30,000 psi)
- 16 voltage inputs (± 10 mv to ± 10 v, 4-20 ma)
- 16 channels of analog output
- 12 pressure outputs (3-15 psig)
- two voltage outputs (0-100 Vdc)
- two rotary positioner outputs (throttle control, etc.)
- Four frequency channels
- 0-500 KHz (programmable time base)
- Rate measurements for speed and flow rates
- Totalizers for mass and volume measurements
- Eight digital inputs and eight digital outputs
- AC or DC (optically isolated)
|
Control and Sequencing
- Control
- 16 PID channels updated at 10 Hz
- Any primary or derived channel can be a process variable
- Eight (opt. 32) digital outputs specified on or off for each
step
- Sequencing
- Up to 340 steps in 64 phases
- Up to 128 profiles per test - each of which can be 512 set
points in length
- Conditional branching between steps
- Step times down to 0.1 second
- Steps can end on time lapse, input, time and input, time or
input, manual entry, or operator intervention
- Process types include: ramp, soak, auto ramp, auto soak, open
loop soak, open loop ramp, manual, profile, and open loop profile
- Alarm and shutdown checks at 1 Hz
- Alarm and shutdown levels programmable per step and channel
|
Signal Processing
- Primary channels (76)
- Configured via user-friendly menus and tables
- All inputs sampled continuously at 10 Hz
- Channels are converted into engineering units
- Thermocouples are linearized
- Each channel digitally filtered at an independent selected
cutoff frequency
- Up to 4th order polynomial linearization
- Polynomial linearization and automatic range selection/detection
for gas analyzers
- Derived channels (32)
- Configured via user-friendly menus and tables
- Calculations can contain 24 built-in functions
- Calculations are updated at a rate of 10 Hz
- Derived channels can be functions of inputs or other calculated
channels
- Each channel digitally filtered at an independent selected
cutoff frequency
- Look-up tables to 1,024 points
|
Data Logging and Display
- Data Logging
- Channels (up to 108) can be logged to disk at one of two
programmable logging rates per phase
- Each phase can have different data logging rates
- Digital filter cutoffs (anti-aliasing) set automatically
- Each channel filter cutoff can be individually specified
- Normal mode logging rates to 10 Hz
- Burst mode to 1 KHz (1-16 channels)
- Display
- Values for up to 48 channels displayed on screen (updated at 1
Hz)
- Position and color of channel display easily selected
- Real-time plots of 8 channels (updated at 1 Hz)
- Plots of snapshots, burst, and stored data
- View operator log file
- Real-time display of PID activity
- Remote real-time display and plot
- Remote date/time set
- Network (Ethernet-FTP) file transfers
|
For more information about UDACS, contact
Robert P. Gauss Jr.,
Principal Engineer, System Design and Instrumentation Section, Fuels and Lubricants Research Division,
Southwest Research Institute, P.O. Drawer 28510, San Antonio, Texas 78228-0510, Phone
(210) 522-3465, Fax (210) 522-5041.
Fuels and Lubricants Research
Brochures
SwRI Brochures
SwRI Publications
SwRI Home
|