|

Optimization of Toxic Gas Sampling From
Fire Effluents and Analysis by Fourier Transform Infrared Spectroscopy, 01-9145
Printer Friendly Version
Principal Investigator
A. Leigh Griffith
Inclusive Dates: 07/01/99 - Current
Background - A large number of fire-related
fatalities result from smoke inhalation rather than exposure to the thermal effects of
fire. For example, a person can be incapacitated by inhaling only 150 to 250 parts per
million of hydrogen cyanide during a five-minute time frame. A system to evaluate the
toxic potency of material in a fire requires three components. First, a fire test
apparatus is needed to generate combustion products by exposing a specimen of the material
to thermal and environmental conditions representative of a real fire. Secondly, a
gas-sampling probe with a pathway from the combustion zone to the measuring device is
necessary to sample the smoke. Thirdly, a calibrated device that can identify and quantify
toxic gases is needed that can analyze the smoke rapidly enough to collect samples
dynamically, thus creating a concentration versus time profile. Many forms of chemical
analysis rely on ex situ grab bag approaches, or are limited to one or two types of gas
species. The use of Fourier transform infrared (FTIR) spectroscopy may result in a single
inexpensive instrument that can perform in-line sampling of fire effluents from both
static and dynamic fire test environments and that can analyze concentrations of multiple
gases quickly with high accuracy. However, while toxic gas analysis based on FTIR
spectroscopy is conceptually straightforward, the practical implementation has several
potential problems. For example, using a high-resolution setting increases noise, making
accurate low-level determinations difficult. Additionally, smoke samples must be
transferred from the test apparatus to the spectrometer at a known temperature and rate to
obtain accurate concentration versus time values. Another problem is that gases are
quantified using a classical least squares fitting of calibration spectra, which does not
account for unknown components in the smoke.
Approach - This program is a systematic study
to optimize the accuracy of toxic gas analysis to a known value from fire tests by
improving gas-sampling techniques, identifying ideal FTIR spectrometer operating
parameters, establishing reliable calibration methods, and developing sophisticated
mathematical techniques for data analysis. The first step in achieving these objectives is
to obtain mixtures of the major gases of interest to establish multipoint calibration
curves of concentration versus absorbance. Because many compounds present a nonlinear
profile, four to five data points will be collected at each range of interest, covering a
range of three orders of magnitude. A gas-mixing system based on partial pressure will be
built to dilute the mixtures to any concentration between zero and the maximum value with
high, known accuracy. The mixer will also blend different gases known to have overlapping
spectral regions that interfere with each other. The second step will involve using these
gas blends in a systematic study of sampling techniques, spectrometer operating
parameters, and mathematical techniques. A survey of different chemometric techniques, for
example, partial least squares, to analyze FTIR spectra will be conducted to determine the
best mathematical technique to approach for this analysis. The third step is to validate
the optimum procedure selected in step two, including all sampling and FTIR parameters and
the quantification technique. This validation will be accomplished by metering known
quantities of a gas into the products of combustion from the fire test apparatus. This
procedure will allow the research team to determine the robustness of the recommended
methods, when used to quantify the composition of real fire gases at high temperature,
with soot, unknown components, and gases in the mixture with overlapping spectra.
Accomplishments - SwRI has constructed a
gas-blending system using a constant-pressure sample cylinder and an ultra-high precision
(0.035 percent) digital pressure gauge. Calibration gases CO, CO2,
HBr, HCl, HCN, HF, NO, NO2, SO2,
acrolein, acetaldehyde, and formaldehyde have been ordered for calibration of the
spectrometer. The survey of chemometric techniques is ongoing. The experimental part of
the program, which consists of calibration, parameter optimization, and validation, is in
progress.
Chemistry and Chemical
Engineering Program
1999 IR&D Home
SwRI Home
|