SwRI: Foaming evaluation procedure is an API CJ-4-approved test used to evaluate foaming-related problems in lubricant oils that can lead to mechanical failure such as inadequate lubrication, cavitation, and overflow loss of lubricant.
| |
 |
| |
The sample, maintained at a
temperature of 24°C
(75°F), is blown with air at a constant rate for
five
minutes, then allowed to settle for 10 minutes. |
|
|
| |
 |
| |
The volume of foam is
measured at the end of
both periods. |
This API CJ-4-approved procedure is used to
evaluate foaming-related problems in lubricant oils. The
tendency of oils to foam can be a serious problem in systems
such as:
-
High-speed gearing
-
High-volume pumping
-
Splash lubrication
This method is used in the
evaluation of oils for operation conditions that lead to
mechanical failure such as:
The foaming procedure method covers the
determination of the foaming characteristics of lubricating
oils at 24°C and 93.5°C. Means of empirically rating the
foaming tendency and the stability of the foam are
described.
Foaming Evaluation Equipment and Procedure
A foaming procedure apparatus, consisting of
a 1000-mL graduated cylinder is fitted with a heavy ring to
overcome the buoyancy, and an air-inlet tube, to the bottom
of which is fastened a gas diffuser. Test baths large enough
to permit the immersion of the cylinder at least to the
900-mL mark and capable of being maintained at temperatures
constant to 0.5°C (1°F) at 24°C (75°F) and 93.5°C (200°F),
respectively. Both bath and bath liquid are clear enough to
permit observation of the graduations on the cylinder. Air
supply from a source capable of maintaining an airflow rate
of 94 ± 5 mL/min through the gas diffuser is required.
Foaming-Related Problems – Evaluation
Method and Summary
The sample, maintained at a temperature
of 24°C (75°F) is blown with air at a constant rate for five
minutes, then allowed to settle for 10 minutes. The volume
of foam is measured at the end of both periods. The test is
repeated on a second sample at 93.5°C (200°F), and then,
after collapsing the foam, at 24°C (75°F).
Pass/Fail Criteria
Foam volumes in mL at the end of three
five-minute blowing period and ten-minute setline period, as
received, and after agitation, are recorded for Sequences I,
II, and III. For the purpose of reporting results, when the
bubble layer fails to completely cover the oil surface and a
patch or eye of clear fluid is visible, the value shall be
reported as “0 foam”.
|
Foaming / Settling |
|
Sequence I, max |
10 / 0 % |
|
Sequence II, max |
20 / 0 % |
|
Sequence III, max |
10 / 0 % |
For more information about the
lubricant oil foaming evaluation capabilities at
Southwest Research Institute (SwRI) or how you can contract with SwRI,
please contact
Becky Grinfield at
bgrinfield@swri.org or (210) 522-3652.
©1998-2009 Copyright Southwest Research Institute
Printed from:
http://www.swri.org/4org/d08/ChemTests/FoamingD/default.htm?
Southwest Research Institute®
(SwRI®) is an independent, nonprofit, applied engineering and physical
sciences research and development organization with
12 technical divisions
using multidisciplinary approaches to problem solving. The Institute occupies
more than
1,200 acres and provides
nearly two million square feet of laboratories, test facilities, workshops, and offices for more than
3,300
employees who perform contract work for industry and government clients.
|